Programming
100 Point Linearization
The FC-5000 flow monitor can be set up to linearize the output from an eligible
flow meter� The calibration data for a particular flow meter and the curve fitted
K-FACTOR and FREQ/VISC points are included when the meter, calibration and
FC-5000 unit are ordered from the factory�
To manipulate or enter the linearization parameters:
1� See "Flow Sensor Type" on page 29 to select Sine UVC Fixed v or Pulse UVC Fixed
v as the sensor input type�
2� Navigate to SYSTEM SETUP > ADVANCED SETUP > SENSOR INPUT >
FLOW SENSOR 1 CAL�
3� Highlight MULTI-POINT TABLE and press ENTEO�
4� For each calibration point, press F2 (f/v) to enter or edit the f/v or F3 (kFct) to
enter or edit the K-factor value�
5� On the numeric entry screens, enter the value and press ENTEO to save the
value, then return to the previous screen�
N
OTEE:
Each entry, 1...100, represents each calibration data point� Any number
of points can be entered, up to 100� Leave the fields at 0�000 if no data
exists�
Flow Sensor Properties
1� Navigate to SYSTEM SETUP > ADVANCED SETUP > SENSOR INPUT�
2� Highlight FLOW SENSOO 1 POOP and press ENTEO�
3� Select and edit each option, as necessary�
The flow sensor properties are:
Option
Ch 1 Bore Dia
Ch 1 Dia� Unit
Ch 1 Material
Ch 1 Smoothing
Page 30
Description
Bore diameter of connected flow meter�
Unit of measure (in� or mm) associated with bore diameter�
Construction material of the connected flow meter�
Smoothing coefficient� As the number increases, averaging
becomes greater� As the number decreases, it approaches the
raw reading�
CTL-UM-01723-EN-01