NOTE:
22
Rechargeable AA batteries will not be recharged in the
instrument. For AA-type rechargeable batteries please use an
external charger.
It is not possible to charge the rechargeable AA batteries or the
RBP2 Li-Ion Battery Pack via the USB interface.
If the PS4 Universal AC/DC Power Supply and the USB interface
are both connected, the instrument is powered by the PS4
Universal AC/DC Power Supply.
Deep discharge
A rechargeable battery that appears to be dead (unit will not
turn on even when connected to external power) may well be in
a deep discharge state. It can be fully charged with the following
charging cycle:
1.
Plug in the power cord for 1–1.5 hours.
Battery should be partly charged now.
2.
Switch the power on.
If the instrument starts: Let the battery fully charge.
– or –
If the instrument does not start:
Unplug power cord, plug it back in and repeat steps 1 and 2.
If the battery is completely deep-discharged, this cycle might
need to be repeated up to 3 times.
General tips on using batteries
• Always handle batteries with care.
• Do not drop or damage the batteries or expose them to
excessively high temperatures.
• Do not store the batteries for more than one or two days at
very high temperatures (e.g. in a vehicle), either separately or
inserted in the instrument.
• Do not leave discharged batteries in the instrument for a long
time if it is not being used.