Appendix E: Auxiliary Electrometer Specifications
Unless otherwise mentioned, all specifications apply at 22C, zero common-mode voltage versus Ground FA,
input voltages with Z
< 10 Ω, and all channel inputs (other than those of the channel under test) at zero volts
out
versus Ground FA. All measurements are made with the Aux Channel A/D with normal Framework calibration.
DC Voltage Measurement
Offset Voltage
Temp Drift
Ranges
Resolution
Gain Error
Offset Range
Offset Resolution
Input Current
Input Impedance
Common Mode
Differential
Common Mode Rejection
CMR Range
CMRR
Crosstalk
Low Frequency
High Frequency
Other AC Specifications
Bandwidth
Phase Shift
Noise
Auxiliary Electrometer Specifications
< 500 µV
< 10 µV/C
50 mV, 500 mV, 5V
1.666 µV, 16.66 µV,
166.6 µV
< 0.3%
5.12 V
166.6 µV/bit
< 10 pA
>30 G in parallel with
<100 pF
>100 G in parallel with
<40 pF
−36 V < Vin < 36 V
> 94 dB
> 86 dB
< −90 dB
< −80 dB
>2 MHz
<1
<1
<4 µV
pp
1/2
<10 nV/Hz
20C to 25C
from 0C to 45C
Full Scale
per bit
zero volts input
−36 V < Vcm < 36 V
−5 V < Vdiff < 5 V
at 3 Hz
at 100 kHz
DC to 10 kHz
10 kHz to 100 kHz
-3 dB
DC to 20 kHz
20 kHz to 100 kHz
0.1 Hz to 10 Hz
at 1 kHz
97
at all gain ranges
Differential Input Voltage
on all ranges
on all ranges
offset of differential signal
measure by voltage
versus floating ground
measured between the
two inputs
versus floating ground