Calibration – Cable Calibration

Procedure to calibrate the cable

1) Connect the Chassis Ground on the back of your potentiostat to a known, good earth ground.
2) Connect the cell cable to the correct color-coded receptacles on the 2 kΩ Calibration Cell.
3) Place the Calibration Cell inside the Calibration Shield, close the lid, and connect the black floating-
ground lead of your cell cable to the Shield's grounding post.