Voltage Measurement

A/D Full-Scale Ranges
Resolution
Zero-Offset Error
Gain Error
Offset Range

Current-to-Voltage Converter

Maximum Full-Scale Range
Minimum Full-Scale Range
Voltage across Rm
Output Voltage (at BNC and ADC in)
Input Offset Current
Range Zero Offset
Gain Tolerance
3000 mA to 3 nA ranges
300 pA range
Zero drift
Bandwidth
3000 mA to 300 µA ranges
30 µA range
3 µA range

Current Measurement

Resolution
Offset Range
Post- Offset Gain
Accuracy
Reference 3000 Specifications
Typ.
12.0, 3.0, 0.3, 0.03
Typ.
400, 100, 10, 1
Max.
1
Max.
0.2
Typ.
10
3000
300
3 (after ×100 gain)
Typ.
150
Typ.
3.0
Max.
5
Typ.
2
Max.
0.05
Max.
0.2
0.5
Typ.
0.03
Typ.
>10
>1.5
>0.15
Typ.
0.00333...
Typ.
 100
Typ.
1×, 10×, 100×
Typ.
Dominated by current-to-voltage error
(see above)
86
Volts
Note 3
µV/bit
mV
Note 11
%
Note 11
V
mA
Note 12
pA
Note 12
pA
mV at
Note 13
full scale
V at
full scale
pA
Note 14
% of
Note 14
range
% of
Note 14
reading
Note 15
% FS/C
MHz
Note 16
% FS/bit
% of
range