(215) 682-9330 9:00 AM–5:00 PM US Eastern Standard Time (877) 367-4267 Toll-free US & Canada Only Replacement parts for this kit are available from Gamry Instruments, Inc. Optional parts may be available from Gamry Instruments, or from third-party vendors. Contact your local Gamry sales representative to discuss any special requirements or accessories that you need.
Table of Contents Introduction ..........................4 Unpacking and Checking a Cell Kit ....................5 Checking for Shipping Damage ..................5 Parts List ........................5 Assembly of the Cell Top ......................6 Important notes ......................6 Procedure ........................6 Requirements for Metal Samples ....................7 Chemical Compatibility .......................
Introduction Introduction The Gamry Instruments’ PTC1 Paint Test Cell kit is a simple, inexpensive cell used for electrochemical corrosion testing. This cell was originally developed by Dr. Richard Granata of Lehigh University. The PTC1 was designed for electrochemical-impedance testing of flat, coated metal specimens. It can also be used for testing corrosion of bare metal samples in highly conductive media.
Gamry part number. If you are checking the completeness of an older kit, most of the components are shown in photographs throughout this manual. Table 1 Parts in Paint Test Cell Kit PTC1 Quantity Part Number...
Assembly of the Cell Top Assembly of the Cell Top This section tells you how to assemble the top of the cell. Warning: The reference and counter electrodes can break during this procedure. If you are not careful, a broken electrode can pierce your hand. Broken glass can have extremely sharp edges that represent a significant safety hazard.
We do not recommend use of the PTC1 with strong acids and bases, oxidizing or reducing agents, or organic solvents. Gamry Instruments is not responsible for damage caused to the cell, your...
Cell Setup Cell Setup 1. Test your thin samples in a setup that includes the PTC1 base (a polypropylene block equipped with four rubber feet). Place a sample on the side of the base without the feet. 2. Place the cell body on top of the sample, making sure that the O-ring fits in the groove on the cell body.
Place the cell body and O-ring of the PTC1 on the PortHole mask so that the exposed area of the sample is enclosed. Clamp the cell body firmly as described in Cell Setup, and proceed normally.
Maintenance Maintenance Replacement Parts All of the part numbers for replacement parts (except for the O-ring) are listed in Table 1. The cell body includes an O-ring. O-rings can also be ordered separately using part number 935-00037 (a package of five O-rings). Reference Electrode Maintenance It is generally best to leave both electrodes in the cell top when the cell is not in use.