Gamry Instruments PTC1 Operator's Manual

Gamry Instruments PTC1 Operator's Manual

Paint test cell kit

Advertisement

Quick Links

PTC1
Paint Test Cell Kit
Operator's Manual
Copyright © 2017 Gamry Instruments, Inc.
Revision 2.1
September 15, 2017
900-00057

Advertisement

Table of Contents
loading

Summary of Contents for Gamry Instruments PTC1

  • Page 1 PTC1 Paint Test Cell Kit Operator’s Manual Copyright © 2017 Gamry Instruments, Inc. Revision 2.1 September 15, 2017 900-00057...
  • Page 3 Limited Warranty Gamry Instruments, Inc. warrants to the original user of this product that it shall be free of defects resulting from faulty manufacture of the product or its components for a period of one year from the original shipment date of your purchase.
  • Page 4 (215) 682-9330 9:00 AM–5:00 PM US Eastern Standard Time (877) 367-4267 Toll-free US & Canada Only Replacement parts for this kit are available from Gamry Instruments, Inc. Optional parts may be available from Gamry Instruments, or from third-party vendors. Contact your local Gamry sales representative to discuss any special requirements or accessories that you need.
  • Page 5: Table Of Contents

    Table of Contents Introduction ..........................4 Unpacking and Checking a Cell Kit ....................5 Checking for Shipping Damage ..................5 Parts List ........................5 Assembly of the Cell Top ......................6 Important notes ......................6 Procedure ........................6 Requirements for Metal Samples ....................7 Chemical Compatibility .......................
  • Page 6: Introduction

    Introduction Introduction The Gamry Instruments’ PTC1 Paint Test Cell kit is a simple, inexpensive cell used for electrochemical corrosion testing. This cell was originally developed by Dr. Richard Granata of Lehigh University. The PTC1 was designed for electrochemical-impedance testing of flat, coated metal specimens. It can also be used for testing corrosion of bare metal samples in highly conductive media.
  • Page 7: Unpacking And Checking A Cell Kit

    Gamry part number. If you are checking the completeness of an older kit, most of the components are shown in photographs throughout this manual. Table 1 Parts in Paint Test Cell Kit PTC1 Quantity Part Number...
  • Page 8: Assembly Of The Cell Top

    Assembly of the Cell Top Assembly of the Cell Top This section tells you how to assemble the top of the cell. Warning: The reference and counter electrodes can break during this procedure. If you are not careful, a broken electrode can pierce your hand. Broken glass can have extremely sharp edges that represent a significant safety hazard.
  • Page 9: Requirements For Metal Samples

    We do not recommend use of the PTC1 with strong acids and bases, oxidizing or reducing agents, or organic solvents. Gamry Instruments is not responsible for damage caused to the cell, your...
  • Page 10: Cell Setup

    Cell Setup Cell Setup 1. Test your thin samples in a setup that includes the PTC1 base (a polypropylene block equipped with four rubber feet). Place a sample on the side of the base without the feet. 2. Place the cell body on top of the sample, making sure that the O-ring fits in the groove on the cell body.
  • Page 11: Porthole™ Electrochemical Sample Masks

    Place the cell body and O-ring of the PTC1 on the PortHole mask so that the exposed area of the sample is enclosed. Clamp the cell body firmly as described in Cell Setup, and proceed normally.
  • Page 12: Maintenance

    Maintenance Maintenance Replacement Parts All of the part numbers for replacement parts (except for the O-ring) are listed in Table 1. The cell body includes an O-ring. O-rings can also be ordered separately using part number 935-00037 (a package of five O-rings). Reference Electrode Maintenance It is generally best to leave both electrodes in the cell top when the cell is not in use.
  • Page 13: Index

    Index Index assembly, 6 base, 7, 8 clamp, 7, 8 counter electrode, 6, 8 damage, 5 distilled water, 10 electrode area, iii, 8 Electroplating Tape 470, 9 KCl, 10 maintenance, iii, 10 O-ring, 5, 7, 8, 9, 10 Parts List, 5 PortHole™, iii, 9 Problems, ii reference electrode, iii, 6, 8, 10...
  • Page 16 734 Louis Drive Warminster PA 18974 Tel: 215 682 9330 Fax: 215 682 9331 www.gamry.com info@gamry.com...