About The Auxiliary Electrometer Option; Notational Conventions - Gamry Instruments Reference 3000 Operator's Manual

Potentiostat/galvanostat/zra
Table of Contents

Advertisement

The Reference 3000 offers an unprecedented combination of high speed, high sensitivity, and low noise. State-
of-the-art analog components were used throughout the design. In all design decisions, performance weighed
more heavily than product cost.
The Reference 3000, like all Gamry potentiostats, requires a computer for its use. Unlike most of Gamry's older
potentiostats, the Reference 3000 connects to the computer through a USB connection. The USB connection
has become truly universal, with USB ports found on all modern computers. Gamry's software currently
supports up to 16 Reference 3000 Potentiostats connected to one computer.
The Reference 3000 is isolated from earth ground. It can therefore be used to make measurements on cells that
contain an earth-grounded metal. A few of examples of such systems include are autoclaves, large metal
storage tanks, stress apparatus, and capillary electrophoresis detectors.

About the Auxiliary Electrometer Option

The Reference 3000 Potentiostat can be equipped with a unique Auxiliary Electrometer option. This factory-
installed option is especially useful when you need to measure the performance of individual cells in a multi-
cell fuel cell or battery stack.
Up to eight completely independent voltages can be measured using this option. The measurements are fully
differential, so cell voltages at any point in a stack can be measured. Each input can measure a 5 V signal
superimposed on a common-mode voltage that can be as large as  36 V! The input impedance is greater than
10
11
, so the inputs can even be connected to small diameter Luggin probes.
This option can be used to simultaneously measure electrochemical impedance on up to eight cells in a cell
stack. This is often of great interest because cells in a fuel cell or battery stack are not identical.
The AE is not restricted to energy-conversion and -storage applications. The electrometer inputs can measure
virtually any voltage. You can measure voltages from temperature, pressure, or strain transducers or voltages of
multiple reference electrodes in a cell.

Notational Conventions

In order to make this manual more readable we have adopted some notational conventions. These are used
throughout this manual and all other Gamry Instruments manuals:
Numbered lists. A numbered list is reserved for step-by-step procedures, with the steps always
performed sequentially.
Bulleted list. The items in a bulleted list, such as this one, are grouped together because they represent
similar items. The order of items in the list is not critical.
Introduction – About the Auxiliary Electrometer Option
16