11. Maintenance
11.5.5. Adsorption Filter Cartridge Maintenance
Observe safety regulations concerning the handling of solvent (see Point 2).
When changing the cartridge, separate the cartridge filter (7) from the economy filter (6) by setting:
Ball valve (1) open
Ball valves (2), (3), (4) closed
Sequence for the rest of the work:
-
Open ball valve (4), drain filter and let it sit and drain sufficiently
(overnight or weekend)
The filter vents automatically over ball valve (5).
-
(The next day): loosen the tension clip on the filter housing and open the housing (7)
-
Unscrew the thumb screw, replace the cartridge (if necessary, replace the
sealing disc) and tighten the thumb screw again
-
Check that the housing gasket sits correctly (replace gasket if necessary).
-
Close the housing, mount the tension clip and then close ball valve (4) on the filter drain.
Fill the cartridge filter with solvent during the next cycle.
Do this by closing ball valve (1) and opening ball valves (2) and (3).
Then either run the filtration over the cartridge filter to the end or
separate the cartridge filter from the economy filter again.
Dispose of used filter cartridges as special waste in a way that avoids emissions!
91
P12 /P15 /P18