Defining Exceedance Triggers - Larson Davis Spartan 730 Reference Manual

Noise dosimeter
Hide thumbs Also See for Spartan 730:
Table of Contents

Advertisement

3.2 Defining Exceedance Triggers

Spartan 730/730IS Dosimeter Manual
Step 4. Highlight, then
list. This opens the Enable screen.
Step 5. Press
up
or
down
Step 6. Press
select
Recommended Next Step:
Select from remaining items in section 3.1 to complete the
measurement configuration.
Exceedance triggers provide a way to monitor for compliance to additional
exposure requirements. For example, if:
workers should not be exposed to C-weighted peak levels above 140 dB
workers' exposure to LAS levels greater than 115 dB should be limited
to a certain duration or to a specific number of occurrences
An exceedance occurs when the measured sound level rises above one or more of
the user-defined trigger levels. For the Spartan 730, there are two threshold levels
(SPL 1 and SPL 2), and three peak trigger levels (Peak1, Peak2, and Peak3). The
process of setting these exceedance levels is explained in section 3.2.1.
The Spartan 730 also allows you to customize the Frequency Weighting, Detector,
and Peak Frequency Weighting for these Exceedance Triggers if you choose to do
so. For the example exposure requirements listed above, you would choose C-
weighting for the peak frequency weighting, and choose S (slow) for the
Detector setting. The process for customizing these settings on the meter is
shown in section 3.2.2.
Defining Exceedance Triggers
select
the firmware option you want to enable from the
to toggle the option to Enabled.
to set the option and return to the Meter menu.
36

Advertisement

Table of Contents
loading

This manual is also suitable for:

Spartan 730is

Table of Contents