Quantitative Analysis
Frame disabling
Frame disabling excludes the actual frame from the cineloop
display. Frame disabling is available only with contrast data.
Disabling frames
1.
To re-enable a
frame: Press
on the cor-
SELECT
2.
responding frame
marker.
disabling successive frames at a time
1.
To re-enable succes-
sive frames: press
and hold down
the
key
SELECT
ECG triggered frame disabling
while dragging the
cursor over the
frame markers.
In a multi-cycle acquisition, the user may deselect all frames in
all heart cycles but a selected one. This function can be used
for example to select a particular systolic frame for each heart
cycle.
1.
2.
3.
4.
5.
Re-enabling all frames
1.
2.
3.
4.
390
Trackball to the frame marker of the frame to disable
beneath the Analysis window (see Figure 8-2).
Press
to disable the frame.
SET
The frame marker turns red.
Press and hold down
frame markers of the frames to disable.
The frame markers turn red.
Scroll through the cineloop to identify the cardiac phase to
analyze or identify the cardiac phase on the ECG trace.
Trackball to the frame marker of the frame of interest in one
of the heart cycles (see Figure 8-2).
Press
.
UPDATE MENU
The System menu is displayed.
Trackball to ECG triggering.
Press
.
SET
All frames in all heart cycles are disabled except for the
selected and corresponding frames in the other heart
cycles.
Trackball the cursor to the Frame marker axis.
Press
UPDATE MENU
panel.
The System menu is displayed.
Trackball to Enable all frames.
Press
.
SET
while dragging the cursor over the
SET
in the trackball area on the control