Getting started
will continue to operate in "Battery Mode", powered by the
internal battery.
While in battery mode, if the AC power is restored within 10
minutes, the system will turn back on within less than 10
seconds.
While in Battery-power mode the system will maintain the
current patient active and not perform "End-Exam". As a result,
when power is restored, the current patient's details and stored
images are readily available on display.
If the AC power is not restored within 10 minutes, the system
will shut-down into Standby mode.
If AC is not restored, the system can remain in Standby mode
for approximately 1 hour using the internal battery power
(assuming battery is fully charged and relatively new). If left for
more than about 1 hour while in standby-mode without AC
power, the system will switch from Standby in to full Shut-down
mode.
Moving and transporting the unit
Wheels
The front wheels of the unit are controlled by the break-pedals
situated above each wheel.
The rear wheels of the unit are controlled by the swivel lock /
release pedals situated above each wheel (see item 10 in
Figure 2-11).
Examine the wheels frequently for defects to avoid breaking or
jamming.
Moving the unit
To prepare the unit to be moved
1.
2.
3.
64
Turn system Off to Full shut-down or Standby mode.
Remove the AC plug from the power outlet.
Wrap the AC power cord to ensure cord is not hanging in
the wheel area or beyond the sides of the system.
Vivid S5/Vivid S6 User Manual