Conditions And Preparations For Surfaces; Mode Of Operation; Switching On And Off - sauter TN-US Instruction Manual

Ultrasonic material thickness gauge
Hide thumbs Also See for TN-US:
Table of Contents

Advertisement

ATU- US 01
2,5
ATU- US 09
5
ATU- US 10
5
ATU- US 02
7
ATB-US01
5
ATB- US 02
5

3.2 Conditions and preparations for surfaces

For any kind of ultrasonic measurement, the condition and roughness of the surface to
be measured is of utmost importance. Rough, uneven surfaces can
restrict the penetration of the ultrasonic waves through the material, resulting in
unstable, incorrect measurement results
The surface to be measured should be clean and free of any substances, rust or
verdigris. If this is the case, the sound generator cannot be cleaned
can be placed on the surface. A wire brush or scraper is often helpful to clean the
surface. In extreme cases, belt sanders or the like can be used. However, it is important
to avoid gouging the surface, which prevents the sound generator from being placed
cleanly.
Extremely rough surfaces like siliceous cast iron are very difficult to measure. These
types of surfaces behave like when light shines on frosted glass, the beam is scattered
and sent in all directions.
In addition, rough surfaces contribute to considerable wear of the sounders, especially
in situations where they are "scrubbed" over the surface.
They should therefore be checked at a certain distance, especially at the first signs of
unevenness on the contact surface. If it is worn down more on one side than the other,
the sound waves can no longer penetrate vertically through the material surface of the
test object. In this case, small irregularities in the material are difficult to measure
because the sound beam is no longer exactly below the sound source.

4. Mode of operation

4.1 Switching on and off

The device is switched on and off by the on/off button.
10
3.0mm~300.0mm
(Stahl)
14
40mm (grey cast
iron HT200)
1.2mm~230.0mm
10
(steel)
1.2mm~230.0mm
10
(Stahl)
0.75mm~80.0mm
6
(Stahl)
0.75mm~80.0mm
6
(in steel)
3~200mm
12
(steel)
20mm
scattering materials
Φ20mm×3.0mm Normal measurement
Normal measurement,
Φ20mm×3.0mm
Φ15mm×2.0mm
Φ15mm×2.0mm
For high temperature
30mm
For thick, highly
damping or highly
90°angle
For thin or slightly
bent tube material
Thin material
measurements
(up to 300°C)
TN_US-BA-e-2020

Advertisement

Table of Contents
loading