10.2
Electrical Specifications
Power supply voltage
Power rating
The input line voltage depends on the local electrical distribution system.
Allowable input voltage fluctuation requirement: ±10 %.
Mode of operation: Continuous operation with intermittent loading
time (at least 60 times the exposure time) before the next exposure begins.
Column operation time: Max. 2 min. On / 18 min. Off (Ratio 1:9)
PaX-i Plus / Insight Installation Manual
Item
Frequency
Accuracy
Description
100 - 240 V ~
50 / 60 Hz
1.3 kVA
Tube Voltage (kVp) ± 10 %
Tube Current (mA) ± 20 %
Exposure Time (s) ± (5 % + 50 ms)