8. Product Specifications
8.3
Electrical Specifications
Power Supply Voltage
Frequency
Power Rating
Tube Voltage
Tube Current
Accuracy Error
The input line voltage depends on the local electrical distribution system.
Allowable input voltage fluctuation requirement: ± 10 %
Mode of operation: Continuous operation with intermittent loading—This
equipment needs a rest time of at least 60 times the exposure time before starting
the next exposure.
Standard: Permanently installed, Option: AC Power Cable
56
Item
Description
100-240 V ~
50/60 Hz (Single phase)
Max. 350 VA
65 kV fixed (± 3 %)
Option A: 2.5 mA
Option B: 3.0 mA
< kVp +10 %, < mA + 20 %, < s ± 5 % or 20 ms
EzRay Air W (Model: VEX-S300W) User Manual