Operation

Setup Procedure

During a calibration process, the calibrator, the product being calibrated,
or it's associated connections, should not to be touched or disturbed. ESD
could cause signal to drop out or lose communication.
1.
Verify host analyzer has software revision A.27.05 or later. On the analyzer press System,
Show System. If an instrument software update is required, the latest revision of the
X-Series signal analyzer software may be downloaded from:
http://www.keysight.com/find/xseries_software
2.
Locate USB cable and Option Upgrade Entitlement Certificate.
3.
Verify Option U9361Y_RCL
The signal analyzer (host) must already have option U9361Y_RCL installed, or the Option
Upgrade Entitlement Certificate supplied can be redeemed to install option RCL. To verify
that option RCL is installed on the signal analyzer select the System > Show System, and
scroll to the bottom of the options listing and look for U9361Y_RCL.
4.
Connect the Module to the Signal Analyzer
Using the USB cable supplied, connect the module to one of the rear panel blue color USB
3 ports. The front panel USB port marked with the lightning bolt can also be used, however
the other two front panel USB ports have insufficient power and cannot be used.
The message Updating Hardware Configuration will appear on screen, and several LEDs on
the module will blink. Then the message USB RCal Connected will appear on screen.
5.
Module Warm-up Time
The Status LED on the module will blink Amber until the module reaches
stability, and then the LED status indicator will turn Green. See
It is recommended that the Receiver Calibrator's Ext Ref IN be connected to either:
The 10MHz Reference OUT port of the Spectrum Analyzer that is controlling the RCAL unit,
or,
The 10MHz Reference OUT port of an instrument that is locked to the 10MHz Reference OUT
of the Spectrum Analyzer that is controlling the RCAL unit.
1. Warm-up time is ~10min (varies with previous conditions: cold start, previously powered-on, etc.)
14
Figure
U9361F/M Receiver Calibrator Quick Start Guide
1
temperature
2.