Keysight Technologies U9361F Quick Start Manual page 18

Receiver calibrator
Hide thumbs Also See for U9361F:
Table of Contents

Advertisement

Example: RCal Calibration Using a UXA Signal Analyzer
Procedure
1.
On the analyzer select: Input/Output > Calibration > Configuration. This should bring up
the RCal Configuration table:
2.
Press Select Calibrator, and replace None with RCal Module 1.
The model number and serial number should now be displayed below the Calibrator option.
These numbers should match the numbers physically on the device.
If RCal Module 1 does not show up as an option, disconnect and reconnect
the USB cable.
3.
Press Insert Row Below to bring up a row in the table.
4.
In the row, select:
Start Freq > 1 GHz
Stop Freq> 3 GHz
Freq Step > 20 MHz
Freq Points > 101
18
U9361F/M Receiver Calibrator Quick Start Guide