Keysight Technologies U9361F Quick Start Manual page 29

Receiver calibrator
Hide thumbs Also See for U9361F:
Table of Contents

Advertisement

Selecting and Positioning the Wrenches
For a few possible connections, the following figures show the proper "flats" on which to place
the wrench.
Ruggedized Adapter to 1 mm Output Port
Figure 9 shows the correct placement of the 10 lb-in torque wrench (the red end) and the 8 mm
open end wrench. On the ruggedized adapter, connect the
8 mm wrench on the flats that are adjacent to the knurled ring.
Figure 9
Wrenches Positioned on Correct Flats
Torque wrench
8710-2819
10 lb-in
torque wrench
end of
8710-2819
U9361M
Output Port
1.0 mm (m)
ruggedized
To connect a cable, or device, to the ruggedized test port adapter, use a 7 mm open end wrench
to hold the test port adapter stationary. For a 2.4 mm cable or device, use a 5/16 inch torque
wrench to torque the cable or device to 8 inch-pounds (0.90 N-m). For a 1 mm cable or device,
use a 6 mm torque wrench to torque the cable or device to 4 inch pounds (0.45 N-m).
U9361F/M Receiver Calibrator Quick Start Guide
Groove marks the correct place
to hold wrench
8 mm open end wrench
1.0 mm
ruggedized connector
Care and Maintenance
Not a connection
(protect from torque)
Do not place wrench
here when installing
ruggedized adapter.
A 7 mm wrench will
be placed here when
connecting a cable.
Ruggedized
adapter
29