Table 33 Lt31 Pwd Status (Byte Number 40); Table 34 Background Luminance Sensor Lm21 (Optional) Alarm Status Information (Byte Number 54) - Vaisala LT31 User Manual

Transmissometer
Table of Contents

Advertisement

Chapter 5 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 173
The byte shown in Table 33 below indicates the status of the PWD
forward scatter sensor which is mounted to the transmitter support unit
LTS111 of the LTT111 (LT31 transmitter).
Table 33
LT31 PWD Status (Byte Number 40)
Code Explanation
0
OK
I
Indication of
abnormal
situation
W
Warning
A
Alarm
E
Error
C
Communication
error
X
Not configured
The byte shown in Table 34 below indicates the status of the LM21
Background Luminance Sensor which is optionally connected to the
receiver interface unit LTI211.
Table 34
Background Luminance Sensor LM21 (optional)
Alarm Status Information (Byte Number 54)
Code Explanation
0
OK
I
Indication of
abnormal
situation
A
Alarm
E
Error
C
Communication
error
X
Not configured
Comment
Cleaning required.
Hardware warning.
Immediate cleaning required.
Hardware error.
Device not responding.
LT31 is not configured to handle
PWD (auto-calibration is off).
Comment
Hood or dew heater problem.
Sensor does not react to changes
in luminance.
LM21 has detected the field
calibrator LMA21.
Backscatter has increased.
Memory error.
Device not responding.
LT31 is not configured to handle
LM21.