9.8 Scaling Function (SYSTEM2)
・ Settings made for one channel can be copied to another channel.
See Section 9.4.
・ When the scaling and waveform processing are simultaneously specified, the
scaling function is not applied to the results of the waveform processing
stored in the internal memory.
(3) Entering the numerical value Method
1. Use the function key to select
2. Use the numerical input key to enter the numerical
value.
Conversion ratio method (conversion ratio: eu/v,
offset)
The setting range: -9.9999E+9 to +9.9999E+9.
2-point method (voltage value: volts, values after
conversion: scale)
The setting range: -9.9999E+29 to +9.9999E+29.
(4) Enter the unit
The unit name can be up to 7 characters long.
1. Move the flashing cursor to
2. Select the
function key. (See Section 9.9.3)
INPUT
If you select List, you can select the unit from the
unit list.
.
use num・key
.