Table 4-10 Register Of Prot Status - Prodigit 3350F Series Operation Manual

High power electronic load
Table of Contents

Advertisement

[STATe:] PROTect?
Syntax: [ STATe:] PROTect?{;NL}
Purpose:Query if there have protection flag which had been set in this 3350F
Description:1.PROT? means the status of Protection of 3350F. "1" means OPP
occurred."4"means OVP. "8" means OCP. Table 4-10 shows the
corresponding number of protection status
2.Use command CLR to clear the register of PROT status to be "0"
bit 7
7 6 5 4 3 2 1 0

Table 4-10 register of PROT status

[STATe:] CCR {AUTOR2}
Syntax: [STATe:] CCR {AUTOR2}{;NL}
Purpose:Set the CC MODE RANGE to be forced to switch to RANGE II
Description:It will switch the RANGE position automatically when setting for AUTO Set
R2 when implementing RANGE II
[STATe:] NGEABLE {ONOFF}
Syntax: [STATe:] NGEABLE {ONOFF} {;NL}
Purpose:To set the GO/NG check function enable or disable.
Description:To set the function of NG judgment opens when POWER ON. When
setting for POWER OFF, the function of NG judgment will not be
implemented.
[STATe:] POLAR {POSNEG}
Syntax: [STATe:] POLAR {POSNEG} {;NL}
Purpose:Set for the display of the voltage meter shows the pole is contrary or not
Description:Set the display of the voltage meter shows the pole. If it shows POS, that
means the pole is not contrary. If the pole is contrary , it will show NEG
[STATe:] START
Syntax: [STATe:]START {;NL}
Purpose:Set for Load to implement the test.
Description : Set for Load to implement the test , and according to TEST
CONFIG(TCONFIG) , the Load will start to test the items and
parameters which are required
[STATe:] STOP
Syntax: [STATe:]STOP {;NL}
Purpose:Set for Load to stop the test
3350F Series Operation Manual 91
bit 0
Over Current Protection (OPP)
Over Temperature Protection (OTP)
Over Voltage Protection (OVP)
Over Current Pttion (OCP)