PRES
Purpose:PRESet ON or OFF status query command.
Command syntax:PRES?{NL}
Description:
PRES? ; return the presently preset status,〝0〞indicates PRESet OFF, and〝1〞indicates
PRESet ON.
MODE
Purpose:CC, CR, CV or CP operating mode query command.
Command syntax:MODE? {NL}
Description:
MODE? ; return the presently operating mode status,〝 0 〞 indicates CC MODE, 〝 1 〞
indicates CR MODE, and〝2〞indicates CV MODE, , and〝3〞indicates CP MODE
CP MODE is available in 33501 electronic load.
NAME
Purpose:33501 Series Electronic Load module model number query command.
Command syntax:NAME?{NL}
Description:
NAME? ; return the activate Electronic Load channel's model number, "33501" indicate the
activate Electronic Load is 33501 Series Electronic load respectively.
PROT
Purpose:OPP, OTP, OVP, and OCP protection status query command.
Command syntax:PROT? {NL}
Description:
PROT? ; return the presently protection status, the status byte register summarizes all of the
protection status events from all status register. Table 4-4 describes the status byte
the happened on the 33501 Series electronic load. The PROT status byte register is
cleared when a CLER command clear all of the PROT and ERR status registers.
bit 7
7
6
5

Table 4-4 PROT status byte register

2
4
3
bit 0
1
0
OPP
OTP
OVP
OCP