Setting Autoranging - Siemens SIPROCESS GA700 Quick Start Manual

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Commissioning: Sequence
8.6 [1.4] Measuring ranges
4. Enable/disable measuring range: "Use measuring range" parameter field.
Note
Using measuring ranges in other device functions/menus
Other device functions or menus can only access measuring ranges activated in this menu.
Inactive measuring ranges are not displayed for selection in other device functions/menus.
This means they can no longer be used, for example, in autoranging or in calibrations.
5. Set additional measuring ranges/parameters, e.g., → Setting Autoranging (Page 106).
8.6.3

Setting Autoranging

Procedure
Note
Autoranging
Autoranging is not possible if the measuring ranges were set incorrectly. In this case measuring
range 1 is selected as active measuring range.
Note
Component-specific setup
Measuring ranges are component-specific. Repeat the procedure described below for all
components, if necessary.
1. "1 Quick Start" > "4. Measuring ranges" > "n component" main menu
The menu includes four navigation lines for setting the measuring ranges and one
navigation line for setting autoranging.
2. Call autoranging: Navigation line "Autoranging".
The parameter display of autoranging has a parameter field and a display box.
3. Specify measuring range: "Measuring range selection" parameter field.
– Authorize access with Standard PIN, if necessary.
– Select one of the offered measuring ranges or autoranging:
106
If the message "Active" is displayed in the "Autoranging" line, autoranging is activated.
If "Not possible" is displayed, correct the start and end values of the individual measuring
ranges: Setting the measuring ranges (Page 104).
Compact Operating Instructions, 05/2015, A5E31805656-03