Combination operation
3.4 Plug & Measure: Permissible activities
Note
Authorization level
To execute the actions described below, you must be logged on with at least the "Expert"
authorization level. You may have to log on again with the requested PIN.
Note
Alarm window
When you change the hardware of the device by replacing or adding modules or components,
the alarm window "Device configuration changed" is displayed in the LUI during re-
commissioning. To start the data synchronization, confirm the alarm by pressing <ENTER>.
Depending on the specific application, parameter sets may have to be transmitted. The
message window "Transmit parameters" is displayed. Information on the further procedure is
available in the tables below:
→ Case differentiations (Page 34)
→ Case differentiations (Page 34)
→ Case differentiations (Page 34)
→ Case differentiations (Page 34)
→ Case differentiations (Page 34)
3.4.3

Case differentiations

Table 3-4
Case
PM
New
New
34
Plug & Measure - Commissioning
Description
AMx
New
Commissioning, PM without serial num‐
ber
New
Commissioning, each module has a se‐
rial number stored at the factory.
None of the modules has been connec‐
ted to another module before
Data synchronization/Action
Automatic data synchronization.
Automatic data synchronization.