Siemens SIPROCESS GA700 Quick Start Manual page 36

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Combination operation
3.4 Plug & Measure: Permissible activities
Table 3-7
Case
PM
Re‐
placed
Re‐
placed
Table 3-8
Case
PM
Un‐
change
d
Un‐
change
d
36
Plug & Measure - Replacement of a PU/Installation of a used AM
Description
AMx
Re‐
Assembly from inventory: Each module
placed
has a serial number and has been con‐
nected to another module before. AM of
the same type is replaced.
Re‐
Assembly from inventory: Each module
placed
has a serial number and has been con‐
nected to another module before. AM of
a different type is replaced.
Plug & Measure - AM slot change
Description
AMx
AM1 →
Internal swapping of modules between
AM2
slots is not permitted!
AM2 →
Install the modules once again in their
AM1
previous slot and restart the device.
AM1 = 0
An AM is removed.
AM2 →
Retrofitting: Remaining AM is inserted
AM1
in the slot that has opened up
Data synchronization/Action
User action. Confirm with <ENTER> or
reject with <ESC>.
When the process is confirmed, stored
AM data is transmitted to the replaced
AM.
Automatic data synchronization, no ac‐
tion required by user. Each module re‐
tains its data.
Data synchronization/Action
None
Automatic data synchronization, no ac‐
tion required by user.
Compact Operating Instructions, 05/2015, A5E31805656-03