Weinmann O2 Box Rescue Description And Instruction Manual page 8

Oxygen supply unit for disaster medicine
Hide thumbs Also See for O2 Box Rescue:
Table of Contents

Advertisement

Available languages

Available languages

Caution!
• Closely observe the instruction manuals and the special information for oxygen
cylinders.
• Secure the oxygen cylinders with cylinder clamps to prevent them from slipping or
falling out.
• When assembling the unit and when changing cylinders, tighten all screwed
unions by hand only. Never use tools for this purpose. Overtightening damages
the screw threads and seals, resulting in leaks.
• Always open the cylinder valve slowly to prevent pressure hammer on the fittings.
• Always ensure that the oxygen cylinders are sufficiently full.
• Do not empty the cylinders completely, as this may allow moist ambient air to
enter them and cause corrosion.
Accessories/Replacement parts
Caution!
Malfunctions and a lack of biocompatibility may result if third-party articles are used.
Please bear in mind that in these cases any warranty entitlement and liability shall
lapse where the accessories recommended in the instruction manual or original
replacement parts are not utilized.
Repairs
Servicing and repair work must only be carried out by the manufacturer (WEINMANN
Emergency) or by a trained member of staff.
Safety information
8

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

Wm 9500

Table of Contents