Weinmann O2 Box Rescue Description And Instruction Manual page 3

Oxygen supply unit for disaster medicine
Hide thumbs Also See for O2 Box Rescue:
Table of Contents

Advertisement

Available languages

Available languages

Legend
1 O
Box Rescue
2
The box contains the oxygen cylinders and the
accessories required for mobile oxygen supply.
2 OXYWAY Click oxygen administrator with
inhalation outlets with locakble dosage
settings
The administrator has four lock adjustable
inhalation outlets. This makes it possible to set the
flow individually for up to four patients. The flow
can be set to 1, 2, 3, 4, 5, 6, 9, 12 or 15 l/min.
Connections can also be made using the Walther
coupler or the DIN coupler.
3 Pressure hose with self-sealing adaptor
and angle connector, 1 m long
The 1 m pressure hose establishes the connection
between the oxygen cylinders and the
administrator.
4 Pressure reducer
The pressure reducers are used to supply medical
oxygen at reduced pressure.
Symbols on the unit label
Symbol
SN
Year manufactured
Serial number of the unit
CE identification according to Medical Device Directive 93/42/EEC
5 Cylinder clamp
They are used to secure the oxygen cylinders in
the O
Box Rescue.
2
6 Oxygen cylinder
Two oxygen cylinders, each holding 5 liters and
with a pressure of 200 bar. The oxygen cylinders
are used to supply medical oxygen.
7 Ventilation mask
This is used to administer oxygen to the patient.
8 Pressure hose with self-sealing adaptor
and DIN connector, 5 m long
The 5 m pressure hose is used:
– to supply the oxygen administrator using a
central gas supply (DIN connection to CGS)
– to supply external medical units (e.g.
MEDUMAT) using the O
connection to administrator)
Meaning
Box Rescue (DIN
2
3

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

Wm 9500

Table of Contents