4. Specification

4.1 Operating Environment

WARNING
atmospheres.
Storage temperatures
Operating temperatures/
environment
ADTS Touch universal
power supply
Power supply
Electrical safety
Operating altitude
Ingress protection
© 2014 General Electric Company – All rights reserved.
This equipment is not rated for use in potentially explosive
ADTS542F: -20°C to 70°C (-4° to 158°F)
ADTS552F )
ADTS553F ) -20°C to 70°C (-4° to 158°F)
ADTS554F )
ADTS542F:
-5°C to 50°C (32° to 122°F)
ADTS552F )
ADTS553F ) -5°C to 50°C (32° to 122°F)
ADTS554F )
Operating humidity: 5% to 95% RH (condensing)
Indoor & Outdoor (non-continuously) use
Operating temperature: 10°C to 50°C (50° to 122°F)
Operating humidity: 5% to 95% RH (non-condensing)
Universal input Mains Power Supply.
Class 1 equipment requiring a protective earth connection.
TXAH 250V HRC Time-delay (T) 250Va.c. rated fuses (X = amp rating
of fuse).
Overvoltage category II, Pollution degree 2
The supplied mains/PSU cable protective earth/ground conductor
must be connected to the power supply protective earth/ground
system.
EN 61010-1, UL 61010-1, CSA 22.2, No. 61010-1 and IEC 61010-1
Maximum 7500ft (2300 metres)
ADTS542/552/553/554 IP33, ADTS Touch IP65 (EN60529)
Operating Environment
ADTS Safety and Installation Guide-English | 15

Advertisement

Table of Contents
loading

This manual is also suitable for:

Druck adts 552fDruck adts 553fDruck adts 554f

Table of Contents