Agilent Technologies G1982-6700 Upgrade Installation Manual page 20

Hplc-chip/nanospray upgrade kit
Table of Contents

Advertisement

11. With the top plate installed, examine the image on the video monitor. If the image is not
clear or appears blurry, the camera is not focused correctly.
12. To refocus the camera, using the supplied 1.5 mm Allen Wrench (part number 8710-2623)
loosen the set (grub) screw that secures the camera in place.
13. In order to focus the camera, examine the image on the monitor while either inserting or
pulling out the camera relative to the HPLC Chip Cube Interface. When the camera is
properly focused, the image on the monitor should appear as below.
14. Once the camera has been focused, tighten the set (grub) screw. Recheck the image on
the video monitor to ensure that the camera has not moved while tightening the set screw.
15. Reinstall the HPLC Chip Cube on the instrument. Be careful not to damage any of the
camera or video cables.
16. Restart the MassHunter Acquisition software if needed to have the HPLC Chip Cube
recognized.
17. Insert a chip into the HPLC Chip Cube. The image on the video monitor should resemble
the image below.
Page 20 of 33
18 July 2011