Assemble the new spray shield and gas diverter assembly
1. While the instrument is tuning or pumping down, locate the following parts:
Winged Nut (part number G1982-20011)
End Cap Nano Spray (G1982-20111)
Radial Gas Diverter (part number G1982-20006)
Wave Spring, Gold Plated (part number G1982-20012)
Contact Spring (part number G1982-20005)
When assembling these parts, wear powder free or lint free gloves. Treat all parts as
ultra clean parts.
2. Start assembling the new spray shield by sliding the Wave Spring (part number G1982-
20012) over the End Cap Nano Spray part (part number G1982-20111).
3. Locate the Radial Gas Diverter (part number G1982-20006) and insert into the End Cap
Nano Spray. Make sure to align the pin in the Radial Gas Diverter with the notch on the
End Cap.