4. Thread the vespel Radial Gas Diverter into Winged Nut (part number G1982-20011). Make
sure to thread the Radial Gas Diverter completely into Winged Nut.
5. Once all of the parts have been assembled, rinse all of these parts with Isopropanol in a
clean, glass beaker at least three times.
6. Once the parts have been rinsed, remove the parts from the beaker and place them on a lint
free cloth.
7. Locate the Contact Spring (part number G1982-20005) and insert it into the cleaned spray
shield assembly.