Replace The Gas Distributor Inside The Desolvation Assembly - Agilent Technologies G1982-6700 Upgrade Installation Manual

Hplc-chip/nanospray upgrade kit
Table of Contents

Advertisement

2. Locate the new Holey Counter Electrode (part number G1982-20008).
3. Using the hex wrench, install the new counter electrode. Set the Holey Counter Electrode
flush against the Electrode Mounting Bracket. Temporarily tighten the set screw to keep
the new counter electrode in place. The final spacing adjustment will be performed with
the interface installed on the instrument.

Replace the gas distributor inside the desolvation assembly

1. Once the instrument has vented, remove the spray shield and the existing endplate. The
stainless steel endplate cannot be used with this upgrade kit. Please dispose it in the
proper disposal container. The new endplate that is part of this upgrade kit is compatible
with all source types.
Please note: If the instrument type is a TOF or Q-TOF and the instrument has the UIRM,
remove the UIRM hardware. Place this hardware in a clean bag as this hardware will no
longer be used with the HPLC-Chip interface. However, it can still be used with the
standard source types supported with the TOF or Q-TOF.
Page 9 of 33
18 July 2011