1. Install the HPLC Chip Interface onto the desolvation assembly. Please note that the Top
Plate should not be installed at this time.
2. Connect the power cables for the illumination LED and video cable for the camera.
3. Locate the Electrode Spacing Tool (part number GT430-20358).
4. Loosen the set screw securing the Holey Counter Electrode. Be careful not to over-loosen
the set screw or the Holey Counter Electrode will fall out of its holder
5. Insert the Electrode Spacing Tool between the counter electrode and the spray shield. The
Electrode Spacing Tool must touch the Holey Counter Electrode AND the new spray shield
assembly. This means the spacing is set correctly. See the figure below as an example.
Page 17 of 33