Agilent Technologies G1982-6700 Upgrade Installation Manual page 10

Hplc-chip/nanospray upgrade kit
Table of Contents

Advertisement

Discard the existing
endplate.
2. Remove the Chamber and Capillary High Voltage cables from the T-nut connectors.
3. Remove the two existing high voltage T-nut connectors that are used to secure the end
plate and discard.
4. Remove the currently installed capillary. A new capillary is supplied as part of this upgrade
kit.
Please Note: This capillary will not be used after installation of this kit. Destroy this
capillary so that it cannot be reused in the instrument or any other instrument.
5. Remove the four T-20 torx screws used to secure the metal ring that holds the Teflon spray
shield support block.
Page 10 of 33
Save the UIRM Union for
use with other sources.