Item
No.
Item Name
————————————————————————————————————————
028
GAS TEMP HILIM
029
REF TEMP #1
030
REF TEMP #2
031
CASE TEMP
032
CASE TEMP MAX
033
CASE TEMP MIN
034
BASE TEMP
035
TCAL ZERO
036
1 PREV TCAL ZERO
037
2 PREV TCAL ZERO
www.honeywell.com
92
Mini-AT User Guide
Description
High limit for temperature, initiates alarm ( items 106 and
146) if exceeded. Also used with Item 004.
Default = 165.00
The temperature value used during the most recent TZERO
calibration.
Default = 0.00
The temperature value used during the most recent TCAL
calibration.
Default = 0.00
Temperature measured inside instrument case at last wake-up.
Default = 0.00
Highest case temperature measured by unit since last manually
reset.
Default = -40.00
Lowest case temperature measured by unit since last manually
reset.
Default = 099999.99
Base temperature per the contract, used to calculate
temperature correction factor (item 045).
Default = 60.00
The most recent offset value for the temperature transducer,
obtained during a T-Zero calibration. The value is normally
very close to 0.0000 and is recalculated during each
Temperature Zero Calibration.
Default = 0.0000
The previous value of item 035, updated on each Temperature
Zero Calibration.
Default = 0.0000
The previous value of item 036, updated on each Temperature
Zero Calibration.
Default = 0.0000