Item
No.
Item Name
————————————————————————————————————————
009
PCOR PRESS MAX
010
PCOR PRES HILIM
011
PCOR PRES LOLIM
012
ATM CALREF
013
BASE PRESS
014
ATM PRESS
015
PCOR REF PRES#1
016
PCOR REF PRESS#2
017
PCOR CAL ZERO
018
1 PREV PCOR CAL ZERO The previous value of item 017, updated on each PCor
www.honeywell.com
90
Page 90
Mini-AT User Guide
Description
Highest PCor pressure measured by the unit since last
manually reset.
High limit for PCor pressure, initiates alarms (items 105 and
145), if exceeded.
Default = 099999.99
Low limit for PCor pressure, initiates alarms (items 105 and
143), if exceeded. Also used with item 003.
Default = 0.00
Local atmospheric pressure entered by the instrument
technician during the most recent calibration of an absolute
pressure transducer.
Default = 0.0000
Base pressure per the contract, used in calculating pressure
correction factor (item 044).
Default = 14.7000
Average atmospheric pressure per the contract, used in
calculating pressure correction factor (item 044) when a gauge
transducer is installed (item 112 = Gauge). This value is
ignored if an absolute transducer is installed.
(item 112 = Absolute).
Default = 14.7000
The pressure value used during the most recent PCor ZERO
calibration.
Default = 0.00
The pressure value used during the most recent PCor SPAN
calibration.
Default = 0.00
The most recent offset value for the pressure transducer,
obtained during a PCor-Zero calibration. The value is
normally very close to 0.0000 and is recalculated during each
PCor Pressure Zero Calibration.
Default = 0.0000
Pressure Zero Calibration.
Default = 0.0000