13. Record the x, y, and RGB luminance values for the center point.
a) Point the spectroradiometer at point 5 (middle square).
b) Click the X value.
c) Record the x value from the spectroradiometer in the projector user interface.
d) Click the Y value.
e) Record the y value from the spectroradiometer.
f)
Click on the Red luminance setting.
The display changes to the appropriate color.
g) Record the luminance value from the spectroradiometer.
h) Click on the Green luminance setting.
The display changes to the appropriate color.
i)
Record the luminance value from the spectroradiometer.
j)
Click on the Blue luminance setting.
The display changes to the appropriate color.
k) Record the luminance value from the spectroradiometer.
l)
Enter the values in MENU > Configuration > Horizontal 1D Color Uniformity.
14. Record the luminance values for the remaining four points (2—inside left, middle square; 15—
far left, middle square; 8—inside right, middle square; and 16—far right, middle square)
a) Point the spectroradiometer at a point.
b) Click on the Red luminance setting.
The display changes to the appropriate color.
c) Record the luminance value from the spectroradiometer.
d) Click on the Green luminance setting.
The display changes to the appropriate color.
e) Record the luminance value from the spectroradiometer.
f)
Click on the Blue luminance setting.
The display changes to the appropriate color.
g) Record the luminance value from the spectroradiometer.
h) Enter the values in MENU > Configuration > Horizontal 1D Color Uniformity.
i)
For the remaining three points, repeat steps a to h.
15. Select MENU > Configuration > Horizontal 1D Color Uniformity > Color Uniformity
Mode.
16. To enable color uniformity, select Horizontal 1D Color Uniformity.
17. Select Enter.
18. To verify 1D color uniformity is working, enable a flat white test pattern.