Checking The Cylinder Pressures (Visual Inspection); Checking The Inlet Pressures (Visual Inspection) - Honeywell ENCAL 3000 Hardware Manual

Gas chromatograph
Table of Contents

Advertisement

8.2.3

Checking the cylinder pressures (visual inspection)

Step
Step
Step
Step
1.
2.
3.
4.
5.
8.2.4

Checking the inlet pressures (visual inspection)

Step
Step
Step
Step
1.
2.
3.
4.
EnCal 3000 Hardware Manual
Action: Reading the gauges
Action: Reading the gauges
Action: Reading the gauges
Action: Reading the gauges
Read the calibration gas cylinder pressure on the pressure gauge.
If there is still a minimum pressure of over 1.0 MPa, the cylinder can remain in
use. Note: if the pressure is close to the minimum, the cylinder must be replaced
independently of the maintenance work if necessary.
Read the carrier gas cylinder pressure on the pressure gauge.
If there is still a minimum pressure of over 1.0 MPa, the cylinder can remain in
use. Note: if the pressure is close to the minimum, the cylinder must be replaced
independently of the maintenance work if necessary.
Please notify your TAC if you require a new gas cylinder. You may use the
device without calibration gas but not without carrier gas. In this case, the
results are less accurate.
Action:
Action:
Action: Reading the gauges
Action:
Reading the gauges
Reading the gauges
Reading the gauges
Read the process gas inlet pressure of the sample gas streams (1 – 5) on the
pressure gauges.
If the inlet pressure (PG) is not like specified (see chapter "Technical
Specifications), adjust the pressure within this range
If you cannot complete step 2, disconnect the gas and voltage supply. Do not
proceed with any further action or work unless step 2 is correct.
Attempt the repair work described in sections 5 and 6 of this manual. If
necessary, notify your TAC.
73022344 o
MAINTENANCE AND REPAIR
93