Followind figure shows the number and position of cable glands (looking from the wall
towards the device). The relevant installation guidelines must be obeyed for routing the
cables. The insulation must at least comply with flammability class 1. Use screened
cables.
Figure 6.9: Position of the cable end glands on the housing base
The electrical inputs are sealed for reasons of explosion protection. In the
event of a replacement, the complete gland must be replaced. The
CAUTION
residual cable in the device must then be cut and can no longer be used.
The replacement and extension of the electrical connections involve similar work to a
new installation. Proceed as described and
in any event, obey the explosion protection regulations.
Use only suitable and approved cable inlet equipment. The cable inlets
must satisfy or exceed the IP rating. NEC-compliant cable inlet
equipment must be used in the USA. CEC-compliant cable inlet
ATTENTION
equipment must be used in Canada.
Cable glands with a sealing system are typically required or, to fulfil the FM regulations,
seal all conduits within 18 inches. See below table for use of gland or conduit
Approval
Glands / Conduits
ATEX
Glands
IECEx
Glands
NEC (US)
Conduit
EnCal 3000 Hardware Manual
Additional info
Use Barrier glands
Use Barrier glands
seal within 18 inches (45 cm)
73022344 o
HARDWARE INSTALLATION