1.5

Additional technical help and repairs

Training classes
The operator must use high-pressure reduction systems and safety
devices to ensure that the maximum inlet pressure, even in the event of a
fault; of the respective gas does not exceed the maximum permitted
value.
The following limits apply:
ATTENTION
Sample gas/Calibration gas/ Verification gas: 0.40 MPa(g) (4.0 bar(g)).
The equipment shall not be used with process gases which contain
oxygen or any other oxidizer in concentrations greater than found in
CAUTION
normal air.
The system operator must ensure that the electrical installation does not
give rise to any danger to personnel or components. The operator must
ensure that no earth potential shift takes place (for example by using an
ATTENTION
equipotential bonding cable).
The operator must ensure that the installation and maintenance work
described in this document is carried out correctly.
The operator is responsible for ensuring that the device remains in
ATTENTION
technically perfect, safe condition throughout its period of operation.
If you have any questions, any points in this document are unclear or you require repair
work, the TAC or our customer service department will be delighted to help. The contact
details are provided at the start of this manual
Honeywell holds technical training classes that are taught by process control systems
experts. For more information about these classes, contact your Honeywell
representative, or see
http://www.automationcollege.com
EnCal 3000 Hardware Manual
ABOUT THIS MANUAL AND DEVICE DELIVERY
Carrier gas: 0.55 MPa(g) (5.5 bar(g)).