HOUSING INTERIOR WITH MAJOR INTERNAL COMPONENTS
5.2.1

Save parameters in the measuring unit

Setting the "Configuration Jumpers" can protect the parameters on the measuring unit
computer witch was loaded by using the software RGC 3000. This configuration Jumper
is at the same position for the all mainboard types. (see following figure) If the jumper is
set, the parameters inside the measuring unit cannot be override. Changes in the
parameters can only be transferred to the measuring unit, if the jumper is not set.
The setting of the jumper, can be checked on the display of the "instrument status" page.
See Section 3.2 of the software manual. How to set the jumper is shown in below.
Figure 5.6: Save parameters in the measuring unit all bord versions
EnCal 3000 Hardware Manual
73022344 o
40