13 Anti-tippers
WARNING
After ANY adjustments, repair or service and BEFORE use, make sure all attaching hardware is tightened securely -
otherwise injury or damage may occur.
Anti-tippers MUST be attached at all times. Inasmuch as the anti-tippers are an option on this wheelchair (you may order it
with or without the anti-tippers), Invacare strongly recommends ordering the anti-tippers as an additional safeguard for the
wheelchair user.
Anti-tippers MUST be fully engaged and release buttons fully protruding out of adjustment holes.
Ensure both anti-tippers are adjusted to the same mounting hole.
13.1 Installing Anti-Tippers
1.
Press release buttons (A) in and insert anti-tippers (B) with the anti-tipper wheels (C) pointing toward ground/floor into the rear frame tubing
(D) until bottom release button locks in place.
2.
Measure the distance between the bottom of the anti-tipper wheels
and the ground/floor.
1
A 1
/
to 2-inch clearance between the bottom of
2
the anti-tipper wheels and the ground/floor MUST
be maintained at all times.
3.
If the distance between the bottom of anti-tipper wheels and the
ground/floor is not 1½ to 2-inches, adjust anti-tippers. Refer to
Adjusting Anti-Tippers on page 92.
Part No 1167417
C
FIGURE 1
91
Anti-tippers
D
A
B
Installing Anti-Tippers