9 Arms
WARNING
After any adjustments, repair or service and before use, make sure all attaching hardware is tightened securely - otherwise
injury or damage may occur.
9.1
Removing/Installing and Adjusting T-Arms
Removing/Installing T-Arms
Install
1.
Align the T-arm (A) with the socket (B) in the T-arm bracket (C).
2.
Lower the T-arm into the T-arm bracket until the height adjustment screw rests on the top of the T-arm bracket.
Remove
1.
Raise the T-arm (A) up and out of the T-arm bracket (C).
Adjusting T-Arm Height
1.
Remove the T-arm. Refer to Removing/Installing T-Arms on page 61.
2.
Remove the height adjustment screw (A) and locknut (B).
CAUTION
When performing STEP 3, DO NOT overtighten adjustment screw, tigthen only to 50 inch pounds of torque. Otherwise,
damage to the product will occur.
3.
Secure the height adjustment screw and locknut in one of the three height adjustment holes (C). Tighten to 50 inch pounds of torque.
4.
Reinstall the T-arm. Refer to Removing/Installing T-Arms on page 61.