Intermediate Cleaning Process - SYSMEX CyFlow Ploidy Analyser Instrument Operating Manual

Hide thumbs Also See for CyFlow Ploidy Analyser:
Table of Contents

Advertisement

Instrument Operating Manual
After finishing the cleaning process CyView
Laser sources will also be switched off.
Switch off the PC
Close Windows (Start – Exit- Switch Computer Off)
To power off completely, use the main power switch at the back panel
5.4

Intermediate cleaning process

The Ploidy Analyser can be cleaned between sets of samples using cleaning solution
(green solution, Order No 04-4009) followed by a sample tube with sheath fluid. This
procedures will allow you to reduce significantly the cross contamination and reduce the
background.
For this procedure, connect a sample tube with cleaning solution and clean the instrument
by clicking on the "Clean" symbol.
The instrument will perform one cleaning cycle. Repeat the procedure with a sample tube
filled with sheath fluid.
Please notice the differences between the clean-modes. If you choose the Clean-symbol
display at right lower corner, the system performs one cleaning cycle. In contrast to the
Clean-symbol display in the upper part, if this symbol will be select the complete
shutdown process will be performed.
Sysmex Partec GmbH
Am Flugplatz 13, 02828 Görlitz, Germany · Phone +49 3581 8746-0 · Fax +49 3581 8746-70 · info@sysmex-partec.com ·
CyFlow® Ploidy Analyser Rev-03_2104-10-22
TM
will be shut down automatically.
www.sysmex-partec.com
26 / 47