Instrument Operating Manual

10.2 Maintenance

Warning: The Waste may contain biohazardous and
carcinogenic material from the samples (infectious
material, dyes).
®
Clean the CyFlow
Water should not enter the CyFlow
contact with electric connections and switches. For cleaning the screen, always use
special screen cleaner and soft cloth.
Do not use any organic solvents, nitro thinner, benzol, alcohol, highly concentrated
bleach etc.!
For cleaning of the flow cuvette, refer to the described cleaning procedure. Do not use
tools to clean the flow cuvette. In case the flow cuvette is blocked, enquire Sysmex Partec
for rapid exchange.
Regularly empty the waste bottle and clean with warm detergent solution and a brush.
Clean sheath reservoir with distilled water and a clean brush and flush with clean distilled
water several times.
Remember: a clean sheath fluid reservoir is critical for proper operation.
®
If the CyFlow
Ploidy Analyser will not be used for longer periods, clean flow system by
using distilled water. Put a sample tube half-ways filled with distilled water at the sample
port. Clean waste and sheath reservoir, wipe top dry.

10.2.3 Service

All service is to be made from an authorized service engineer. Please contact your local
supplier or Sysmex Partec (support@sysmex-partec.com).
For further information, please consult our web site:
Products
Application tips
Instrument and software support
News and events
And much more ...
www.sysmex-partec.com
Sysmex Partec GmbH
Am Flugplatz 13, 02828 Görlitz, Germany · Phone +49 3581 8746-0 · Fax +49 3581 8746-70 · info@sysmex-partec.com ·
Ploidy Analyser casing on a regular base carefully with soft cloth.
®
Ploidy Analyser or peripheral devices or come into
CyFlow® Ploidy Analyser Rev-03_2104-10-22