RESULTS

4.1.2
The RESULTS register defines properties of calculated results displayed in the RESULTS
table.
Properties of result calculations
It is possible to set up calculations with the COUNT of individual regions according to the
specified formula:
NumReg1 (+ - x / ) NumReg2
--------------------------------------------------x Scale
DenomReg1 (+ - x / ) DenomReg2
Result will be display in the "Results table" next to the "Region statistics"
Sysmex Partec GmbH
Am Flugplatz 13, 02828 Görlitz, Germany · Phone +49 3581 8746-0 · Fax +49 3581 8746-70 · info@sysmex-partec.com ·
NumRegion1/2 defines numerators
DenomRegion1/2 defines denominators
NumOperator defines operator between 2
numerators „+", „-", „x" or „/"
DenomOperator defines operator between 2
denominators „+", „-", „x" or „/"
Unit allows to add text to the result table
Scale introduces a factor to the formula
Counter results on refers to the result of a
Volumetric counting
Create a result by pressing "New" or delete a
result by pressing "Delete"
Confirm all modifications by pressing Accept