Chapter 9 Cleaning And Maintenance - SYSMEX XE-5000 Instruction Manual

Automated hematology analyzer
Hide thumbs Also See for XE-5000:
Table of Contents

Advertisement

9. Cleaning and Maintenance
Daily maintenance
Periodical SRV maintenance
As-needed maintenance
Supplies replacements
Adjustment of pressure and vacuum
Sysmex XE-5000 Instructions for Use
To ensure that the instrument can function in its best state, it
is necessary to give scheduled maintenance.
Perform maintenance according to the schedule below and
record the result in the maintenance checklist.
• Execute shutdown (detector chamber and dilution line are
cleaned automatically).
• Remove fluid from the Pneumatic Unit trap chamber.
• Clean the sample rotor valve when the message "Clean the
SRV" is displayed:
Depending on the service contract, your Sysmex technical
engineer will perform the cleaning of the SRV.
• Clean the manual rinse cup.
• Clean the sample rotor valve tray.
• Clean the piercer tray.
• Removing Clogs. (Clog Removal Sequence)
• Clean the IMI detector aperture.
• Clean the RBC detector aperture.
• Remove flow cell air bubbles in the optical detector block.
• Clean the flow cell in the optical detector block.
• Replace the waste container. (option)
• Replace reagents.
• Replace piercer.
• Replace hand clipper.
• Replace the rubber plate No. 39.
• Replace the HEPA filter.
• Replace fuses.
• Adjust the pressure to 0.25 MPa.
• Adjust the pressure to 0.16 MPa.
• Adjust the pressure to 0.07 MPa.
• Adjust the pressure to 0.03 MPa.
• Pneumatic Unit vacuum
Stop Bar
• Vacuum: Adjust the vacuum to -0.04 MPa.
Before operating with the front cover open, be sure to
set the stop bar. Otherwise, the cover can drop down
and injure your head.

CHAPTER 9 Cleaning and Maintenance

Warning!
9-1