Daily Cleaning And Disinfecting; Packaging Symbols - SYSMEX XW-100 Instructions For Use Manual

Automated hematology analyzer
Hide thumbs Also See for XW-100:
Table of Contents

Advertisement

2. Clean spills from the surface with a new Clorox Wipe. Wipe down the surface area with
a different new Clorox Wipe to disinfect.
3. Allow the disinfected areas to sit for at least one minute. Wipe down all disinfected areas
with a towelette (or paper towel) moistened with water.
4. Dispose of all cleaning materials in a biohazard container. Allow surfaces to air dry prior
to use.
5. Remove gloves and thoroughly wash hands with soap and water after disinfection.

Daily Cleaning and Disinfecting

1. Put on personal protective equipment; disposable gloves, eye protection, and lab coat.
2. Using a new Clorox Wipe clean and disinfect the system outer housing, display, exterior
of the sample chamber and the white sample adapter used for testing on the XW-100
every day. Wipe down all surface areas with a different new Clorox Wipe to disinfect.
3. Allow the disinfected areas to sit for at least one minute. Wipe down all disinfected areas
with a towelette (or paper towel) moistened with water.
4. Dispose of all cleaning materials in a biohazard container. Allow surfaces to air dry prior
to use.
5. Remove gloves and thoroughly wash hands with soap and water after disinfection of the
device.
6. Daily cleaning and disinfection must be documented. Please refer to Appendix A for a
template to document the completion of this activity.
Caution:
The following are signs of deterioration of the device and/or screen covers. If present, please
contact Technical Assistance.
• Paint and/or plastic flaking
• Rust
• Hazy screen covers
• Quality control is out of range
• Device lock up

Packaging Symbols

This way up
Stacking limit by number
Keep away from rain
Fragile; handle with care
Page 12 of 32