B - Performance and Calibration Procedures

Changing the Calibration Password

The factory default password is 0. You can change the password when the dc source is in calibration
mode (which requires you to enter the existing password). Proceed as follows:
Action
1.
Reset the unit.
2.
Enable calibration mode. (0 is the default password)
3.
Enter the new password. You can use any number with up to six
digits and an optional decimal point. If you want the calibration
function to operate without requiring any password, change the
password to 0 (zero).
4.
Save the password.
5.
Exit Calibration mode. (*RST also exits calibration mode)

Calibration Error Messages

Errors that can occur during calibration are shown in the following table.
Error
401
CAL switch prevents calibration (call the factory for details)
402
CAL password is incorrect
403
CAL not enabled
404
Computed readback cal constants are incorrect
405
Computed programming cal constants are incorrect
406
Incorrect sequence of calibration commands
102
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table B-3. GPIB Calibration Error Messages
Meaning
Program Commands
"*RST"
"CAL:STAT ON, 0"
"CAL:PASS <password>"
"CAL:SAVE"
"CAL:STAT OFF"