Turn-On Checkout

Front Panel Description

1
2
Q Q Q Q
AC mains power switch.
Line
Switch
R R R R
Unit indicators light to indicate the following operating conditions:
Unit
Power The dc source is turned on.
Indicators
Active The dc source is addressed to talk or listen.
Error
S S S S
Channel indicators light to indicate the following channel conditions:
Channel
On
Indicators
Prot
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
N3280A 10V, 0.5A
Component Test DC Source
3
Figure 3-1. Front Panel, Overall View
There is a message in the SCPI error queue.
The specified output channel is enabled.
The specified output channel has entered protection mode due to:
Overtemperature,
Overvoltage,
Oscillation protect, or
Power clear.
Query the status registers of the affected channel to determine which
protection feature is tripped.