[SOURce:]VOLTage:MODE

This command determines what happens to the output voltage during a triggered event.
FIXed
The output voltage is unaffected when a trigger occurs.
The output voltage is programmed to the value set by VOLT:TRIG when a trigger occurs.
STEP
Command Syntax [SOURce:]VOLTage:MODE <mode>, (@<channel list>)
Parameters FIXed | STEP
*RST Value FIXed
Examples VOLT:MODE FIX,(@1)
Query Syntax [SOURce:]VOLTage:MODE? (@<channel list>)
Returned Parameters <CRD>
Related Commands VOLT:TRIG

[SOURce:]VOLTage:PROTection:STATe

This command enables or disables the overvoltage protection (OVP) function. The command only
applies in voltage priority mode. When enabled, the output of the unit will shut down and the output
relays will open when the output voltage exceeds +11.5V (±0.3V), or −11.5V (±0.3V).
CAUTION:
Disabling the overvoltage protection function may cause excessive output voltages, such
as can occur if remote sense leads are shorted, to damage the equipment under test.
Command Syntax [SOURce:]VOLTage:PROTection:STATe <bool>, (@<channel list>)
Parameters 0 | OFF | 1 | ON
*RST Value ON
Examples VOLT:PROT:STAT 0,(@1)
Query Syntax [SOURce:]VOLTage:PROTection:STATe? (@<channel list>)
Returned Parameters <NR1>0 or 1